Equipment Description
33
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Power Entry
Module
Figure 10, Device, left side, power entry module
he power entry module contains:
On/Off switch:
Powers the device on and off. When switched off, all
device functions stop (see
Fuse panel:
Interrupts power to the device in case of excessive current
draw. The specification information regarding the fuse is as follows:
100 - 120 Volts AC: 5 A, 250 V, Type F, 5 mm x 20 mm
200 - 240 Volts AC: 2.5 A, 250 V, Type F, 5 mm x 20 mm
Input power receptacle:
Allows the connection to the power cord.
Electrical line filter:
Reduces the effects of power surge on the device
operation.