78
Preventive Maintenance and Installation
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
11.Press
NO
to bypass the line sensor calibration. The screen displays:
12.Lock the centrifuge cover and press
YES
(the device continues beeping)
to continue. The screen displays:
The centrifuge starts. Once the centrifuge has stopped the screen
displays:
The device continues beeping.
Note: If the bowl makes noise replace it with a new bowl and close and lock the
centrifuge cover.
BOWL SEATING TEST
IS CENTRIFUGE COVER LOCKED
PRESS YES TO CONTINUE
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
BOWL SEATING TEST
TESTING CENTRIFUGE
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
BOWL SEATING TEST
DID BOWL PRODUCE UNUSUAL NOISE
PRESS YES OR NO TO CONTINUE
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
Содержание ACP 215
Страница 10: ......
Страница 38: ......
Страница 58: ......
Страница 104: ...104 Preventive Maintenance and Installation P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual...
Страница 126: ......
Страница 150: ......
Страница 182: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 182 System Schematics System Schematic...
Страница 183: ...Haemonetics ACP 215 Service Manual P N SM ACP215 01 EN AB System Schematics 183 Family Tree...
Страница 184: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 184 System Schematics Pneumatic Schematic...
Страница 186: ......