112
Calibration and Configuration
P/N: SM-ACP215-01-EN(AB)
Haemonetics
®
ACP-215
®
Service Manual
9. Press
START
. The screen displays:
Figure 31, Market protocol configuration
10.Press
MODIFY PROGRAM
to move the pointer to the desired protocol
options that will be available to the operator of the machine.
11.Press
START
. The screen displays:
Note: If the Cell Wash protocol is selected above, only the Cell Wash
configuration would be displayed in the menu below. If the Frozen Blood
protocol is selected above the glycerolization and deglycerolization protocols
would only be displayed below.
12.Press
START
. The screen displays:
Figure 32, Additive option in cell wash
Note: This option is not available in the North American market configuration.
INTERNATIONAL MARKET CONFIGURATION
AVAILABLE PROTOCOLS
-> Cell Wash Only
Frozen Blood Only
Both
Status = Both
PROTOCOL CONFIGURATION
International
-> Cell Wash
Glycerolization
De-Glycerolization
CELL WASH CONFIGURATION
International
-> Additive
Содержание ACP 215
Страница 10: ......
Страница 38: ......
Страница 58: ......
Страница 104: ...104 Preventive Maintenance and Installation P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual...
Страница 126: ......
Страница 150: ......
Страница 182: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 182 System Schematics System Schematic...
Страница 183: ...Haemonetics ACP 215 Service Manual P N SM ACP215 01 EN AB System Schematics 183 Family Tree...
Страница 184: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 184 System Schematics Pneumatic Schematic...
Страница 186: ......