Disassembly
147
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Power Entry Module
Line
Conditioner
1. Remove back panel from device.
2. Unplug the L (brown) and N (blue) AC power lines from the power
supply.
3. Remove the (2) Phillips head screws securing the line conditioner to the
top shelf of the power supply cage.
Power Entry
Module
1. Remove the front and rear panels.
2. Remove (2) 1/4” nuts on either side of the power entry module.
3. Disconnect left top (white or blue) and bottom (black or brown) AC power
lines from the power entry module and disconnect the ground lug from
the ground bus on the cabinet.
Fuse
1. Pry above the power switch on the power entry module to open the fuse
cover.
2. Pry out each fuse holder.
Note: Do Not discard the fuse holder when replacing fuses. Remove the fuse
from the holder and dispose of the fuse only.
3. Replace with same size and rating fuse (5A 250V for 120VAC devices
and 2.5A 250V for 240VAC devices).
Note: When replacing fuse holders, be sure the arrows on the fuse holder and
power entry module face the same direction.
Содержание ACP 215
Страница 10: ......
Страница 38: ......
Страница 58: ......
Страница 104: ...104 Preventive Maintenance and Installation P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual...
Страница 126: ......
Страница 150: ......
Страница 182: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 182 System Schematics System Schematic...
Страница 183: ...Haemonetics ACP 215 Service Manual P N SM ACP215 01 EN AB System Schematics 183 Family Tree...
Страница 184: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 184 System Schematics Pneumatic Schematic...
Страница 186: ......