CALIBRATION
107
S/C
trimming
SHORT THE TEST LEADS
DC
Spot freq
F 1
F 2
F 6
Abort
SELECT A CHOICE
TO CONTINUE...
F 3
F 4
F 5
All freq
<= 100kHz
<= 10kHz
4.
Always select F5 (All freq) when
using the standard fixture set
included in LCR-8101 package.
F 5
DC
(S/C trimming only) Trims at 0Hz.
Spot
freq
Trims at the frequency set in the
measurement mode (page30).
<=
10kHz
Frequency range is 0Hz ~ 10kHz.
<=
100kHz
Frequency range is 0Hz ~ 100kHz.
All
freq
Frequency range is 0Hz ~ 1MHz.
Frequency limit example: When using a special
fixture set, trimming failed at 50kHz which is
outside of the test component range (up to
5kHz). In this case, use F3 (<= 10kHz).
5.
The trimming starts and ends automatically.
SHORT CIRCUIT TRIM
IN PROGRESS
ALL FREQUENCY
LCR-8101
User
Manual
108
Trim
pass
The display goes back to the
calibration mode menu.
Trim
fail
The Fail message appears on the
display. Press any other key to go
back to the original menu.
Press any key to continue...
FAILED
SHORT CIRCUIT TRIM
6.
Switch the fixture setting from O/C to S/C (or
from S/C to O/C) and repeat from step 1.