GRAPH
MODE
83
1.5kH
Enter
H
k
Backspace
All clear
Clear
Increase
Decrease
If a wrong unit is entered, the value is cancelled.
Unit Mismatched
If a value outside of the range is entered, the
closest available value is automatically selected.
Nearest Available
If the entered start Voltage level is higher than
the stop Voltage, the two values are swapped.
Hi and Lo Swapped
Set Lo level
6.
Press the Up/Down key to move the
cursor to Lo level and repeat the
same step.
Cs Lo:
3.5626mF
LCR-8101
User
Manual
84
Set vertical axis ( Percentage)
Background
The Y (vertical) axis configuration is available for:
•
Manual or Automatic fit: Selects whether the
vertical range is manually set or automatically
adjusted as the graph is plotted.
•
Absolute or Percentage: Selects how the vertical
range is defined, as absolute values (minimum
and maximum) or percentage of the nominal
(center) value.
Panel operation
1.
Press F2 to select % (Percentage).
Abs
%
F 2
2.
Press F3 to select Manual Fit.
Manual Fit
F 3
3.
The Hi percentage, Lo percentage, and the
nominal value of the vertical axis appear.
Step Size:
Cs Hi
Cs Lo
Nominal:
:
:
2 4 8
1
-19.500%
10.500%
100.00mF
Set Hi level
4.
Press the Up/Down key to move the
cursor to Hi percentage.
Cs Hi:
10.500%
5.
Use the numerical keys to enter the percentage.
Range
–1.0x10
12
(Tera) ~ 1.0x10
12
(Tera) %
50%
Enter
1200%
k
Enter