GRAPH
MODE
79
If a wrong unit is entered, the value is cancelled.
Unit Mismatched
If a value outside of the range is entered, the
closest available value is automatically selected.
Nearest Available
If the entered start Voltage level is higher than
the stop Voltage, the two values are swapped.
Hi and Lo Swapped
Set stop Voltage
4.
Repeat the above step for the stop Voltage.
Stop:
1.00 V
Range
50mV ~ 2V, 1mV resolution (stop
Voltage must be higher than the start
Voltage)
Set measurement
Frequency
5.
Use the Up/Down key to move the
cursor to Frequency setting (Freq).
Freq :
10.000kHz
Use the numerical keys to enter the
measurement frequency.
Range
20Hz ~ 1MHz
50Hz
Enter
1MHz
M
Enter
Backspace
All clear
Clear
Increase
Decrease
Select Lin/Log
scale
6.
Press F1 to select the horizontal
scale, Linear or Logarithmic.
Log(V)
Lin(V)
F 1
LCR-8101
User
Manual
80
Set horizontal axis (Frequency)
Background
The X (horizontal) axis is selectable from Voltage
and Frequency sweep.
•
When Voltage sweep is selected: measurement
Frequency is fixed
•
When Frequency sweep is selected:
measurement Voltage is fixed
Select Frequency
1.
Press the Up/Down key and move
the cursor to Sweep.
Sweep:
Drive Level
2.
If necessary, press the Left/Right
key to change the sweep setting to
Voltage (Drive Level).
Drive Level
Frequency
Set start
Frequency
3.
Press the Up/Down key and move
the cursor to Start.
Start:
:
20.000Hz
Use the numerical keys to enter the
starting Frequency.
Range
20Hz ~ 1MHz (start Frequency must
be higher than the stop Frequency)
50Hz
Enter
1.2kHz
k
Enter
Backspace
All clear
Clear
Increase
Decrease