REMOTE
CONTROL
95
Interface Configuration
Configure RS-232C interface
Connector
DB-9, Male
RS-232C
configuration
Baud rate
9600
Parity
None
Data
bit
8
Stop
bit
1
Connect the RS-232C cable to
the rear panel port: DB-9 male
connector.
RS232
Pin assignment
1 2 3 4 5
6 7 8 9
2: RxD (Receive data)
3: TxD (Transmit data)
5: GND
7: RTS (Request to send)
8: CTS (Clear to send)
4, 6, 9: No connection
PC connection
Use the Null Modem connection as in the below
diagram.
LCR-8101
PC
TxD
RxD
GND
Pin2
Pin3
Pin5
RxD
TxD
GND
Pin2
Pin3
Pin5
CTS
RTS
Pin7
Pin8
CTS
RTS Pin7
Pin8
LCR-8101
User
Manual
96
Configure GPIB interface
Connection
Connect the GPIB cable to the
rear panel port: 24-pin female
connector.
Pin assignment
12
1
24
13
Pin1
Data line 1
Pin13
Data line 5
Pin2
Data line 2
Pin14
Data line 6
Pin3
Data line 3
Pin15
Data line 7
Pin4
Data line 4
Pin16
Data line 8
Pin5
EOI
Pin17
REN
Pin6
DAV
Pin18
Ground
Pin7
NRFD
Pin19
Ground
Pin8
NDAC
Pin20
Ground
Pin9
IFC
Pin21
Ground
Pin10
SRQ
Pin22
Ground
Pin11
ATN
Pin23
Ground
Pin12
Shield (screen)
Pin24
Signal ground
GPIB constraints
•
Maximum 15 devices altogether, 20m cable
length, 2m between each device
•
Unique address assigned to each device
•
At least 2/3 of the devices turned On
•
No loop or parallel connection