REMOTE
CONTROL
103
:multi:delay
<NR2>
:multi:delay?
Sets or returns trigger delay time for currently
selected step in millisecond.
Parameter: 0ms ~ 1000ms
Example: :multi:delay 10m (10ms)
:multi:freq <NR2>
:multi:freq?
Sets or returns the frequency for the currently
selected step in Hz.
Parameter: 20 ~ 1MHz
Example: :multi:freq 1e3 (1kHz)
:multi:func
<NR1>
:multi:func?
Sets or returns measurement type for the currently
selected step.
Send parameter: LS, LP, Q, CS, CP, D, Z, PHASE,
RS, RP, X, G, B, Y, RDC
Return parameter: 1 (Z), 2 (Ls), 3 (Lp), 4 (Cs), 5
(Cp), 7 (Y), 8 (G), 9 (P), 10 (Q), 11 (D), 12 (Rs), 13
(Rp), 14 (B), 15 (X), 16 (Rdc)
Example: :multi:func ls (Series inductance)
:multi:hi-lim
<NR3>
:multi:hi-lim?
Sets or returns the higher test limit of the currently
selected step.
Example: :multi:hi-lim 10 (limit to 10.0)
:multi:lev <NR3>
:multi:lev?
Sets or returns the drive level for the currently
selected step in Voltage.
Parameter: 10mV ~ 2V
Example: :multi:lev 200m (200mV)
:multi:load
<filename>
Loads an existed file to run or edit.
Example: :multi:load demo (file name demo)
:multi:lo-lim
<NR3>
:multi:lo-lim?
Sets or returns the lower test limit of the currently
selected step.
Example: :multi:lo-lim –5 (limit to –5)
:multi:new
<filename>
Create a new multi-step program.
Example: :multi:new demo (file name demo)
LCR-8101
User
Manual
104
:multi:res?
Query the results of the test for each step.
Parameter: 0 (Pass), 1 (Fail Hi), 2 (Fail Lo)
Example: 1, +1.5E-7, 0, –0.2E-4 (step 1 failed on
high limit, step 2 passed)
:multi:run
Switches to the multi-step run page.
:multi:save
Save currently edited file.
:multi:speed
<disc>
:multi:speed?
Sets or returns the measurement speed for the
currently selected step.
Send parameter: Max, Fast, Med, Slow
Return parameter: 0(Max), 1(Fast), 2(Med), 3(Slow)
Example: :multi:speed max (maximum speed)
:multi:test <NR1>
:multi:test?
Selects or returns the step being edited.
Parameter: 1 ~ 30
Example: :multi:test 1(step 1 selected)
:multi:trig
Starts running multi-step measurements.
Calibration command
:cal:oc-trim
<NR1>
Performs open circuit trimming.
Parameter: 1 (Spot trim), 2 (<10kHz), 3 (<100kHz),
4 (All frequency)
Example: :cal:oc-trim 4 (calibrate for all frequency)
:cal:sc-trim
<NR1>
Performs short circuit trimming.
Parameter: 1 (Spot trim), 2 (<10kHz), 3 (<100kHz),
4 (All frequency), 5 (Rdc)
Example: :cal:sc-trim 4 (calibrate for all frequency)
:cal:res?
Returns the result of the calibration performed.
Parameter: 0 (fail), 1 (pass)