Command Set
System command
*cls
Clears the Event Status Register and associated
status data structure.
*ese <NR1>
*ese?
Sets or returns the current contents of the Standard
Event Status Enable Register as an integer in the range
0 to 255.
*esr?
Returns the current contents of the Standard Event
Status Register as an integer in the range 0 to 255.
It also clears ESR.
*idn?
Returns oscilloscope ID as Manufacturer, Model
No, Serial No, Firmware version.
Example: GW INSTEK, 8101, 0, 1.84
*loc
Sets the instrument to local state.
*opc
Sets the OPC bit of the ESR register.
*opc?
Always returns 1 as instrument commands are
always processed sequentially.
*opt?
Returns the hardware options installed in the
instrument.
Example: 1MHz, GPIB, RS232, GRAPH MODE
*rst
Resets LCR-8101.
*sre <NR1>
*sre?
Sets or returns the current contents of the Service
Request Enable Register as an integer in the range 0 to
63 and 128 to 255.
:stat:oper:con?
Reads Status Operation Condition register.
:stat:oper:enab
<NR1>
Sets Status Operation Enable register.
:stat:oper:even?
Reads Status Operation Event register.
LCR-8101
User
Manual
100
*stb?
Returns the current contents of the Status Byte
with the Master Summary bits as an integer in the
range 0 to 255. Bit 6 represents Master Summary
Status rather than Request Service.
*trg
Triggers a direct measurement, but does not return
the results to the controller. This is the same as a
GET (Group Execute Trigger) command.
*wai
Command has no effect as commands are
processed sequentially.
Measurement command
:dump-bmp
Returns the display as a windows compatible
bitmap.
:beep <disc>
:beep?
Sets or returns the buzzer condition.
Set parameter: off (disabled), pass (beeps when
passed), fail (beeps when failed)
Return parameter: 0 (off), 1 (pass), 2 (fail)
:loc-trig <NR1>
Turns On/Off local triggering in remote control
Parameter: on (local control), off (remote control)
:meas:equ-cct
<NR1>
:meas:equ-cct?
Selects or returns equivalent circuit.
Send parameter: ser, par
Return parameter: 0 (parallel), 1 (series)
:meas:freq
<NR3>
:meas:freq?
Sets or returns frequency of AC measurement in
Hz.
Parameter example: (1kHz) 1k, 1000 Hz, 1E3
:meas:func
<disc>
Selects first or second AC measurement function.
Parameter: c, l, x, b, z, y, q, d, r, g
Example: :meas:func:c;d (C+D measurement)
:meas:func:major
?
Returns the first AC function.
Parameter: 0 (C), 1 (L), 2 (X), 3 (B), 4 (Z), 5 (Y)