PASS-FAIL
MODE
65
Edit program step
How to edit
parameter
•
For selecting parameters, press F1
(Prog) repeatedly.
F 1
•
For entering values, use the numerical and unit
keys.
Example:
0.5kHz
k
Enter
Backspace
All clear
Clear
Increase
Decrease
How to move
cursor
To move the cursor to the editing point,
use the arrow keys: up, down, left, and
right.
Note: The cursor cannot move into
blank columns or steps.
Panel operation
1. Create new
step
To create a new step in a program,
press F1 (Prog) at Func column “OFF”.
The step becomes active and the
function changes to Ls. Up to 30 steps
are available in a program.
F 1
Step
Func
OFF
01
Step
Func
Ls
01
2. Select item
(function)
Move the cursor to Func column, press
F1 (Prog) repeatedly. The measurement
item (function) changes in the following
order.
F 1
Ls
→
Lp
→
Q
→
Cs
→
Cp
→
D
→
Z
→
θ
→
Rs
→
Rp
→
X
→
G
→
B
→
Y
→
Rdc
→
Ls
LCR-8101
User
Manual
66
Set frequency
Move the cursor to Freq column.
Enter the frequency using the
numerical keys and unit keys.
Func
Freq
500.00
Ls
2.00 V
Volt
Range
20Hz ~ 1MHz, 5 digit resolution
Example:
0.5kHz (500Hz)
k
Enter
Set voltage
Move the cursor to Volt column.
Enter the voltage using the
numerical keys and unit keys.
Freq
Volt
2.00 V
500.00
Bias
Range
10mV ~ 2V, 1mV resolution
Example:
100mV
m
V/A
Enter
Select data
capture rate
Move the cursor to Spd column.
Select the capture rate by
pressing F1 (Prog) repeatedly.
Bias
Spd
MAX
1.0000H
Hi
F 1
Slow
500mS
Med
450mS (AC), 120mS (Rdc)
Fast
150mS (AC), 50mS (Rdc)
Max
75mS (AC), 30mS (Rdc)
Set Hi limit
Move the cursor to Hi column.
Enter the Hi limit using the
numerical keys and unit keys.
Spd
Hi
1.0000H
0.0000H
Lo
MAX
Range
follows the specification for each
measurement item
Example:
1.5kH
(for Ls)
Enter
H
k