BASIC
MEASUREMENT
43
Show circuit model or scale (pass/fail)
Background
The center of the display can be filled with the
diagram of equivalent circuit model, or the
measurement scale with Pass/Fail test result. This
selects not only the diagram/scale but also
whether running the Pass/Fail test or just
measuring the value.
Panel operation
Press F3 (Show/Hide scale) to select
circuit model or scale.
F 3
Normal
MEASUREMENT MODE
0.01234
0.2179
mF
D
*7
C
L X B Z Y
Q
D R G
Parallel
Show Scale
50mVac
1.5000kHz
Range
Speed Med
Auto
Vm:549.8mV
Im:724.9pA
–
Pass/Fail test
MEASUREMENT MODE
0.01234
0.2179
mF
D
*7
C
L X B Z Y
Q
D R G
Parallel
Hide Scale
Abs
%
Save Nom
HI
Hi
Lo
20.0mF
0.00mF
50mVac
1.5000kHz
Range
Speed Med
Auto
Vm:549.8mV
Im:724.9pA
–
Pass/Fail test
For Pass/Fail test details, see page52.
LCR-8101
User
Manual
44
Parameter Configuration
Select measurement item
*This is not necessary for Rdc measurement.
Measurement
combination
The following list shows the available combination
of the first and second measurement items.
Capacitance (C) Series
C-Q, C-D, C-R
Parallel C-Q, C-D, C-R, C-G
Inductance (L) Series
L-Q, L-D, L-R
Parallel L-Q, L-D, L-R, L-G
Reactance (X)
Series
X-Q, X-D, X-R
Susceptance (B) Parallel B-Q, B-D, B-R, B-G
Impedance
(Z)
Z-Angle
Admittance
(Y)
Y-Angle
Panel operation
To select the first measurement
item, press F1 repeatedly.
C
L X B Z Y
F 1
To select the second measurement
item, press F2 repeatedly.
Q
D R G
F 2
To select the circuit model, series or
parallel, press F3 repeatedly.
F 3
Series
Parallel