PASS-FAIL
MODE
67
Set Lo limit
Move the cursor to Lo column.
Enter the Lo limit using the
numerical keys and unit keys.
Spd
Hi
0.0000H
Lo
MAX
1.0000H
Range
follows the specification for each
measurement item
Example:
1.0kH
(for Ls)
Enter
H
k
Set trigger delay
Move the cursor to Dly (Delay)
column. Enter the amount of
trigger delay time using the
numerical keys and unit keys.
Hi
Lo
10 ms
Dly
1.0000H
0.0000H
Range
0 (no delay) ~ 1000ms
Example:
10ms
m
LCR-8101
User
Manual
68
Copy (duplicate) program step
Background
Copying the step inserts a new, identical step next
to the current step (= the step where the cursor
resides).
Panel operation
Press F2 (Copy). A new step with
identical contents appears on the right
side.
F 2
Before (step 3 empty)
After (step 2 copied to 3)
Step
Func
Freq
Volt
Bias
Spd
Hi
Lo
Dly
500.00mS
01
02
03
1.0000k
10mV
MAX
FAST
1.0000 S
9999 ms
0 mS
1.00 V
0.0000
Ω
0.0000
Ω
Rdc
B
FAST
0 mS
1.00 V
0.0000
Ω
0.0000
Ω
Rdc
Delete program step
Background
Deleting the step deleted the currently selected
step (= where the cursor resides). The other step
numbers are decreased by 1 (shifted to the left in
the table).
Panel operation
Press F3 (Delete). The current step is
deleted, and the whole table shifts to
the left.
F 3
Before (step 2 deleted)
After (step 3 becomes 2)
Step
Func
Freq
Volt
Bias
Spd
Hi
Lo
Dly
500.00mS
01
02
03
OFF
1.0000k
10mV
MAX
FAST
1.0000 S
9999 ms
0 mS
1.00 V
0.0000
Ω
0.0000
Ω
Rdc
B