REMOTE
CONTROL
101
:meas:func:minor
?
Returns the second AC function.
Parameter: 0 (Q), 1 (D), 2 (R), 3 (G)
If the first function is Z or Y, this command returns
the last non-polar setting
:meas:hi-lim
<NR2>
:meas:hi-lim?
Sets or returns scale high limit as percentage.
Example: :meas:hi-lim 5.0 (+5.0%)
:meas:lev <NR2>
:meas:lev?
Sets or returns drive level for currently selected
test.
Parameter example: (200mV) 0.2V, 200m
:meas:limit
<disc>
:meas:limit?
Sets or returns percentage or absolute scale limits.
Send parameter: abs (absolute), perc (percentage)
Return parameter: 0 (absolute), 1 (percentage)
:meas:lo-lim
<NR2>
:meas:lo-lim?
Sets or returns scale low limit as percentage.
Example: :meas:hi-lim –5.0 (–5.0%)
:meas:nom
<NR3>
:meas:nom?
Sets or returns nominal value for scale.
Send parameter: according to the active unit (1e-6f
= 1uF)
Return parameter example: .10000000e-1 = 10mH
:meas:range
<NR1>
:meas:range?
Selects or returns auto-ranging or range-hold on
range N.
Send parameter: auto, hold, 1 ~ 7
Return parameter: 0 (auto), 1 ~ 7
:meas:scale
<disc>
:meas:scale?
Shows or hides the scale bar or returns the status.
Send parameter: on, off
Return parameter: 0 (scale hidden), 1 (scale visible)
LCR-8101
User
Manual
102
:meas:speed
<disc>
:meas:speed?
Selects or returns measurement speed.
Send parameter: max, fast, med, slow
Return parameter: 0 (max), 1 (fast), 2 (med), 3
(slow)
:meas:test:ac
Selects AC measurement.
:meas:test:rdc
Selects Rdc measurement.
:meas:test?
Returns measurement type.
Parameter: 0 (AC measurement), 1 (Rdc
measurement)
:meas:trig
Triggers an AC or Rdc measurement manually.
Returns the 1
st
and 2
nd
measurement (only the 1
st
in
Rdc).
Example: –396.283E-6, 99.558 (uF/D)
:mode?
Query the currently selected operating mode.
:rep <disc>
:rep?
Enables or returns repetitive measurements when
unit is returned to local control.
Send parameter: on (repetitive), off (single shot)
Return parameter: 0 (single shot), 1 (repetitive)
Example: :rep on (repetitive mode)
:trig
Triggers a measurement in the current mode.
Multi-step program command
:multi:set
Switches to the multi-step set-up page.
:multi:del
Removes a step in the program.
Parameter: 1 ~ 30
Example: :multi:del 2 (deletes step 2)