PASS-FAIL
MODE
53
Single-Step Test Configuration
Overview
Background / test
type
Pass/Fail test checks whether the measurement
result sits between hi and low limit. Two methods
are available: absolute limit and percentage limit.
Absolute
limit
The Hi and Low limit are defined by
absolute values.
Low
Limit
Hi
Limit
Pass
Percentage
limit
The Hi and Low limit are defined by
the distance (percentage) from the
Nominal value.
Nominal
– ##%
Pass
Nominal
Nominal
+ ##%
C
S
Series capacitance
X
Reactance
Test item
C
P
Parallel capacitance
B
Susceptance
L
S
Series inductance
Z
Impedance
L
P
Parallel inductance
Y
Admittance
R
S
Series resistance
R
DC
DC Resistance
R
P
Parallel resistance
θ
Angle
For detailed description of each item, see page31.
LCR-8101
User
Manual
54
Configure beep setting
Background
The beep sounds when the pass/fail test result
matches the setting: failed or passed.
Panel operation
1.
Press the Menu key, then F5
(System). The system configuration
appears.
Menu
F 5
Precision LCR Meter LCR-8101
5
:
Frequency
Software version 2.00 Jan 20 2007
RS-232
GPIB
Graph mode
Line frequency
Beep
GPIB address
:
:
OFF
50Hz
1MHz
2.
Press the Up/Down key and move
the cursor to Beep.
Beep
:
OFF
3.
Press the Left/Right key to select the
beep setting, Off, Pass, or Fail.
Off
Beep is turned Off
Pass
Beeps when the test result is pass
Fail
Beeps when the test result is fail
Beep in repetitive
mode
If the repetitive measurement is On, the beep
might sound continuously. If this becomes a
problem, either use the Single mode (press
Sing/Rep key) or turn Off the beep.