Multi-Step Program Run
Run program
Panel operation
1.
When editing is completed, press F6
(Run) to run the multi-step program.
The display changes to program
running mode.
F 6
MULTI STEP MODE
F 1
PROGRAM: Demo
Run
–
Start
F 6
SET
2.
Press the Sing/Rep key to select
Single (manual trigger) or Repetitive
(automatic trigger).
Sing/Rep
Manual
trigger
Manual trigger
Starts running the program
when the Trigger key or F1
(Start) is pressed.
Trig
F 1
Auto
trigger
Auto trigger
Starts running the program when
LCR-8101 detects the DUT is
connected to the fixture (scans the
fixture continuously). Manual
triggering is also available.
LCR-8101
User
Manual
70
3.
Press F1 (Start) or the Trigger key to
manually start the program. The test
results show up according to the
program contents.
Trig
F 1
MULTI STEP MODE
PROGRAM: Demo
Run
–
Start
SET
1
2
3
4
Freq
1.2000k
10.000k
100.00k
DC
Volt
1.00
1.00
1.00
1.00
Ls
Q
Ls
Rdc
Result
9.8936mH
22.708 Q
10.852mH
25.555
Ω
LO
PASS
HI
PASS
The rightmost row shows the result for each
step.
Failed: below the Lo limit
Failed: above the Hi limit
Passed
The left bottom corner shows the result for the
whole program.
PASS
All steps passed
FAIL
One or more steps failed
4.
To return to the program setting
menu, press F6 (Set).