89
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
Measurement modes
Description
Flicker Mode
This mode performs voltage fluctuation and flicker measurement in
accordance with IEC61000-3-3 and IEC61000-4-15 international
standards.
FFT Mode
This mode can show the power spectrum of the input signal through
FFT (Fast Fourier Transform). Please use this mode to view the
frequency details of an input signal.
Cycle Mode
This mode can measure voltage, current, power and other
parameters of each AC input signal cycle.
8.5 Measurement Items
Table 8-17: Measurement items
Items
Symbols and Meanings
Voltage(V)
Urms: TRMS, Umn: ARV calibrated to
RMS
Udc: Linear Average Voltage, Urmn:
Rectified Mean Voltage
Uac: Voltage RMS that remove DC signal
Support for simultaneous
measurements,
peak
factor up to 300
Current(A)
Irms: TRMS, Imn: ARV calibrated to RMS
Idc: Linear Average Current, Irmn:
Rectified Mean Current
Iac: Current RMS that remove DC signal
Support for simultaneous
measurements, peak
factor up to 300
Active Power(W)
P
Apparent Power(VA)
S
Reactive Power(Var)
Q
Power Factor
λ
Phase Difference(°)
φ
Frequency(Hz)
fU(FreqU): voltage frequency, fI(FreqI): current frequency
Maximum and
Minimum Voltage (V)
U+pk: maximum voltage, U-pk: minimum voltage
Maximum and
Minimum Current (A)
I+pk: maximum current, I-pk: minimum current
Peak Factor
CfU voltage peak factor, CfI current peak factor
Corrected Power (W)
Pc (applicable standards: IEC76-1(1976), IEEE C57.12.90-1993,
IEC76-1(1993))
Efficiency
Measurement of efficiency
η