PA5000 Series Power Analyzer Start Guide
Figure 4-2: Wiring system selection
(1) Operation procedure
Step 1: Wiring key to enter the wiring configuration menu.
Step 2: Select the
“Wiring Setting” button, turn the menu operation knob and select
the input element of the wiring system, as shown in Figure 4-2.
2. Wiring method
In order to ensure measurement accuracy, the stray capacitance and the influence
factor of measuring voltage and current must be considered.
Influence of stray capacitance
When measuring a single-phase device, the current input terminal of the instrument is
connected to the terminal of Source near the ground. This reduces influence of stray
capacitance on the measurement accuracy, as shown in Figure 4-3.
Figure 4-3: Reduce influence of stray capacitance on the measurement accuracy
Influence of measuring voltage and current on the measurement
Note the position of input terminals according to the measuring voltage and current,
as shown in Figure 4-4.