85
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
500kHz~1MHz.
1000V
330V
33V
3.3V
10Hz 45Hz
10kHz
100kHz
500kHz
1MHz
Frequency
Voltage
Current accuracy
Within the range of 10Hz
~
45Hz, if the current exceeds 3A, the current
accuracy is just reference value.
Within the range of 400Hz
~
1MHz, if the current exceeds 20A, the current
accuracy is just reference value.
Within the range of 5kHz
~
1MHz, if the current exceeds 3A, the current
accuracy is just reference value.
Within the range of 10kHz
~
1MHz, if the current exceeds 0.33A, the current
accuracy is just reference value.
The current accuracy is just reference value within the range of
30kHz~1MHz.
50A
3A
0.33A
10Hz 45Hz
10kHz
100kHz
500kHz
Frequency
Current
400Hz
30kHz
20A
5kHz
For the waveform display data, the accuracy of Upk and Ipk should be their
respective accuracy value + 3% of the range (reference value). However, for the
external sensor, its accuracy should be added 3% of input accuracy +5mV
(reference value). Valid input range is within
±
300% of the range.
DC measurement additional error
DC voltage accuracy value is added 0.5mV; power accuracy value is added
0.5mV × current reading.
External sensor input current accuracy value is added
50μV, power accuracy
value is added
(50μV/ sensor range) of the range × 100%.