86
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
5A element direct input current accuracy value is added
20μA, power accuracy
value is added
20μA × voltage reading.
50A element direct input current accuracy value is added 1mA, power accuracy
value is added 1mA × voltage reading.
Temperature variation
Voltage DC accuracy value is added 0.3mV/
℃
of the range.
The DC accuracy of the current input should be added the following values.
5A input element: 20
μA /
℃
50A input element: 1mA /
℃
The DC accuracy of the external current sensor input is added
50μV/
℃
.
DC power accuracy: additional voltage error × additional current error.
The self-heating effect of the input signal (U is the voltage (V), I is the current (A).
If the input voltage AC exceeds 400Vrms, the voltage and power accuracy
should be added 10
-7
×U
2
% of the reading.
If the input voltage DC exceeds 400Vrms, the voltage and power accuracy
should be added 10
-7
×U
2
% of the reading and plus 10
-7
×U
2
% of the range.
5A input element current and power accuracy
Input AC current: 6×10
-3
I
2
% of the reading.
Input DC current: 6×10
-3
I
2
% + 4×10
-3
I
2
μA of the reading
50A input element current and power accuracy
Input AC current: 6×10
-5
I
2
% of the reading
Input DC current: 6×10
-5
I
2
% + 4×10
-3
I
2
mA of the reading
Even if the input current decreases, the self-heating effect will continue until the
temperature of the internal shunt resistor drops.
The impact of data update rate on accuracy
All accuracy values must be added 0.5% of reading when the data update rate is
10ms.
All accuracy values must be added 0.1% of reading when the data update rate is
50ms.
All accuracy values must be added 0.05% of reading when the data update rate
is 100ms.
The impact of small range on accuracy
5A element 10mA range: I
rms
is in the allowable range of 1~110%; the current and
power accuracy is added 10mA×reading (mA)
-1
× 0.02% of the range.
5A element 20mA range: I
rms
is in the allowable range of 1~110%; the current and
power accuracy is added 20mA×reading (mA)
-1
× 0.003% of the range.
50A element 1A range: I
rms
is in the allowable range of 1~110%; the current and
power accuracy is added 1A×reading (A)
-1
× 0.005% of the range.