PA5000 Series Power Analyzer Start Guide
8.2 Monitor
Table 8-13: Display parameters
8.3 Accuracy
8.3.1
Basic Accuracy
The measurement accuracy of power analyzer is obtained under the following
conditions:
Temperature: 23±5°C. Humidity: 30~75%RH. Input Waveform: sine wave. Common
Mode Voltage: 0V. Line Filter: OFF.
λ (Power Factor): 1. Crest Factor: 3. Warm-up time
of 30 minutes. Perform zero calibration before measurements. f = frequency. Data
update Rate: 500ms.
Table 8-14: Measurement accuracy of PA5000 series 5A power element
Indicator ± (% r%measurement range)
Frequency range of
input signal
Current/Voltage/Sensor
Power
DC
0.05+0.10
0.05+0.10
0.1Hz
≤ f <
30Hz
0.10 + 0.10
0.20+ 0.20
30Hz
≤ f <
45Hz
0.10 + 0.10
0.10 + 0.20
45Hz
≤ f <
66Hz
0.10 + 0.05
0.10 + 0.05
66Hz
≤ f <
1kHz
0.10 + 0.10
0.20 + 0.10
1kHz
≤ f <
10kHz
0.15 + 0.10
0.30 + 0.10
10kHz
≤ f <
50kHz
0.20 + 0.10
0.30 + 0.20
50kHz
≤ f <
100kHz
0.50 + 0.30
0.70 + 0.50
100kHz
≤ f <
500kHz
(0.004f+0.8) + 0.5
(0.02f-0.3) + 1.0
500kHz
≤ f ≤
1MHz
(0.01f-2.2) + 1.0
(0.042f-12) + 2.0
Display Parameters
Parameters Description
Monitor
12.1
” color LCD display
Resolution
1280×800 pixels
Touch Screen
Support the touch screen operation
Display Update Rate
Same as the data update rate