PA5000 Series Power Analyzer Start Guide
Figure 2-17: Cycle-by-cycle analysis view
2.7.9
Normal Analysis Mode
In some applications, the measuring instrument should save the measured data, and
perform data measurement and analysis operation on the instrument; In addition, in
some applications, the measurement data changing quickly should be saved first,
and then called out to perform analysis one by one.
In the normal mode, the power analyzer will automatically save the current
measurement data, and then enter into the normal analysis mode. Users can recall
the current saved measurement data as well as perform all measurement and
analysis operations in the normal mode.
2.7.10
Waveform Operation
In addition to the waveform display, the waveform operation module can also use the
customized formula to perform waveform operation and display them. The power
analyzer can set two sets of formulas at the same time. Each formula can use the U, I
as variants to perform operation. The commonly used functions such as Sin, Cos and
Sqrt are included in the formula.