57
©2016 Guangzhou ZHIYUAN Electronics
Date: 2016/12/08
User Manual
V1.00
C
Guangzhou ZHIYUAN Electronics Co., Ltd.
PA5000 Series Power Analyzer Start Guide
Figure 5-2: Measurement menu
5.3.2
Steps
1. Overview
Users can enable or disable the auto-zeroing function and perform manual zero
calibration operation in the measurement menu.
2. Enter into measurement menu
Press the
“Measure” button in the Function Key area on the front panel to enter into
measurement menu, as shown in Figure 5-2.
3. Enable/Disable auto-zeroing function
Press the
“auto-zeroing” soft key in the measurement menu shown in Figure 5-2, and
then users can select whether to enable the auto-zeroing function in the pop-up dialog
box.
4. Perform manual zero calibration
Press the
“zero calibration” soft key in the measurement menu shown in Figure 5-2 to
perform manual zero calibration. For accurate measurements, it is recommended that
the instrument should be warmed up for more than 30 minutes before performing zero
calibration.
5.4 Configuration Wizard
The power analyzer provides a configuration wizard for users. It helps users to
complete the basic configuration of instrument before starting measurement. Press the