![GS Chromalox 4081 Скачать руководство пользователя страница 183](http://html1.mh-extra.com/html/gs/chromalox-4081/chromalox-4081_installation-and-operation-manual_2274774183.webp)
175
Command Code High Byte
68
44
Command Code Low Byte
80 or 65
50 or 41
Delete A Profile (DP) = 80dec / 0x50hex
Delete All Profiles (DA) = 65dec / 0x41hex
Profile Number High Byte
A/R
A/R
Note: The profile number is not included
in the message when deleting all profiles.
Profile Number Low Byte
A/R
A/R
CRC High Byte
A/R
A/R
CRC Low Byte
A/R
A/R
The instrument replies to this message with an Edit Response Message.
The delete segment command deletes the specified segment from the specified profile. The following segments
are moved up one place in the profile (e.g. if segment 6 is deleted segment 7 becomes segment 6).
Delete a Segment - Request (to instrument)
Field Name
Data
Comments
Dec
Hex
Unit Address
A/R
A/R
The network address ID of the instrument.
Function Code
23
17
Requires the multi read/write function.
Read Start Address High Byte
32
20
Read Start Address Low Byte
6
6
Read Quantity Of Registers High
0
0
Read Quantity Of Registers Low
1
1
Write Start Address High
32
20
Write Start Address Low
6
6
Write Quantity Of Registers High
0
0
Write Quantity Of Registers Low
02 or 01
02 or 01
Byte Count
04 or 02
04 or 02
Command Code High Byte
68
44
Command Code Low Byte
83
53
Profile Number High Byte
A/R
A/R
Profile Number Low Byte
A/R
A/R
Profile Number High Byte
A/R
A/R
Profile Number Low Byte
A/R
A/R
CRC High Byte
A/R
A/R
CRC Low Byte
A/R
A/R
The instrument replies to this message with an Edit Response Message.
Deleting a Segment