Reverse osmosis system
GENO-OSMO RO 125K
Order no. 285 752 940-inter Edited by: phar-mrie G:\BA-752940-INTER_285_RO_125K.DOCX
40
Display /
factory
setting
Parameters
Setting
range
Comments
5. 0
Mains return reaction
for fault Er 0 (mains
failure > 5 minutes).
0 ... 2
0 = Irrespective of
whether the system
was switched off or
on before the mains
failure, it remains
switched off after the
mains returns and
fault Er 0 is output.
1 = Fault Er 0 is
deactivated.
2 = After the mains
power returns, the
system is switched
off or on as it was
before the mains
failure, and fault Er 0
is output.
6. 1
Daily interval for forced
operation/forced
withdrawal [days].
1 ... 3
If the day interval since
the last permeate output
has been reached at the
programmed time, forced
operation or forced
withdrawal takes place –
depending on what is
activated.
For RO125K, only
forced withdrawal is
permitted in
conjunction with the
solenoid valve forced
withdrawal option!
7.18:00
Time forced operation/
forced withdrawal.
00:00 ...
23:59
8. 0
Duration forced
operation [hours].
0 ... 9
9. 3
.
0
Opening time solenoid
valve forced
withdrawal [minutes].
0.0 ...
99.9
A. 0
Recovery monitoring
(Er 5).
0 … 1
With AVRO 125 TS/TL
the recovery monitoring
must be activated!
b. 80
Upper recovery limit
value [%].
1 … 99
Caution!
With
AVRO 125 TS/TL,
the recovery must
be set to 50 %!
c. 060
Delay time for recovery
deactivation [min.].
0 … 240