Reverse osmosis system
GENO-OSMO RO 125K
Order no. 285 752 940-inter Edited by: phar-mrie G:\BA-752940-INTER_285_RO_125K.DOCX
36
Key functions:
Basic function:
Expanded function in programming levels:
Acknowledgement of malfunctions.
Access to time programming (press
and hold button > 2.5 s).
Open parameter for editing (value is flashing).
Save and close parameter.
Switch off system (> 5 s in basic time
display)
Decrease numerical value.
Return to previous menu item.
Switch on system (> 5 s in basic time
display), display operating values of
the information level
Increase numerical value.
Go to the next menu item.
+
Access to the code-protected
programming levels (code request
C 000)
+
Close the opened parameters without saving
(previous value is maintained).
Return to basic display time.
2.1
How to read the
operating state
Various operating parameters can be displayed in the information level.
The information level is accessed by pressing the ▲ button (> 5 s).
The other parameters are accessed by touching. The information level
remains locked for as long as the system has not been enabled using
the close input signal.
Key
Display
Parameters
00:00
Basic time display
The system might still be switched on with the first press (> 5 s)!
365
Remaining duration of the service interval [days]
0000
Display value not relevant for RO 125 K TL/TS
LF022
Permeate conductivity [µS/cm] (optional –
display value flashes if the advance warning
value is exceeded)
P0200
Permeate flow rate [l/h]
c0200
Concentrate flow [l/h]
u0320
Flow rate blending [l/h] (option 752 800)
A 050
System recovery [%]