Reverse osmosis system
GENO-OSMO RO 125K
Order no.
285 752 940-inter Edited by: phar-mrie G:\BA-752940-INTER_285_RO_125K.DOCX
31
1.2 Rinsing out
preserving agent
For the duration of storage and transport, the membrane is protected
by means of a preserving agent. First of all, this preserving agent
must be flushed out. In order to prevent the system being switched on
beforehand, it is locked electronically.
Note:
For more detailed information on the handling of the control unit,
refer to chapter F.
Via code 113, (refer to Table E-1: Extract from…), parameter ECL
release the system: To do this, open parameters with the P key,
use the ▼ key to set ECL: 1 and confirm with the P button
Via code 113, (refer to Table E-1: Extract from…), parameter
EnL: 1, open both solenoid valves ("DEAERATE") and rinse
preserving agent out system for 30 minute by opening parameters
with the P button, use the ▲ button to set EnL: 1 and confirm with
the P button
Complete program step "DEARATE": Open parameter with
P button, set EnL: 0 with ▼ button and confirm with P button
Exit the "EnL" program by simultaneously pressing the ▼and
▲ buttons