CS610-UM-251-9370
8-16
8. DISASSEMBLING AND ADJUSTING THE MECHANICAL PARTS
8.12 Contact Glass Assembly
Detaching the contact glass assembly
(1) Open the center cover.
(2) Remove the six M2L5 binding head screws holding the front contact glass stopper plate.
M2L5 binding head screw
M2L5 binding head screw
M2L5 binding head screw
M2L5 binding head screw
(3) Detach the front contact glass stopper plate from the unit.
(4) Remove the six M2L5 binding head screws holding the rear contact glass stopper plate.
(5) Detach the rear contact glass stopper plate from the unit.
(6) Slide the contact glass assembly to the right side, lift up the contact glass from the left side, and then
detach the contact glass assembly from the unit.
Reattaching the contact glass assembly
(1) Reattach the contact glass in the reverse order in which it was detached.
(2) Make sure that each of the CIS assembly surfaces lies flat against the inside of the contact glass
assembly when you reattach the contact glass assembly.
(3) Perform the software adjustment.
When replacing the contact glass assembly, be sure to clean the interior of the image sensor unit and the inside of
the contact glass assembly. A soiled or dusty unit can affect scanning performance.
Содержание CS610-11eN
Страница 1: ...COLORIMAGESCANNER SERVICE MANUAL CS610 SERIES CS610 UM251 02 9370...
Страница 2: ......
Страница 42: ......
Страница 44: ......
Страница 118: ......
Страница 120: ......
Страница 123: ...CS610 UM 251 9370 13 3 13 PARTS LIST Main Frame 1 7 6 4 Ethernet board Enhanced I F board 8 9 3 5 2...
Страница 128: ...13 PARTS LIST CS610 UM 251 9370 13 8 CIS Unit 6 4 2 11 9 10 12 8 1 3 5 7...
Страница 130: ...13 PARTS LIST CS610 UM 251 9370 13 10 Top Cover 3 5 2 9 6 10 7 4 8 8 10 11 12 12 12 13 14 15 16 1...