PART THREE - OPERATION
SECTION 8 - HART OPTION
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
86
8.5 HART Preferences
Setup
Changing
Polling Address
Viewing Number of
Required Preambles
Use a hand-held HART terminal or HART-capable PC to set
HART preference information. When using a Model 275
HART Communicator to access preference menus, select
the “GLI SETUP” line in the MAIN MENU screen and press
the
key
to reveal this screen:
M33/53C: GLI
GLI SETUP
1 HART INFO
sdd sdsdsds
2 DEVICE INFO
3 MASTER RESET
SAVE HOME
Use the “HART INFO” submenu to:
•
Change the polling address used by the master to
identify a device (analyzer).
•
View the number of preambles required by a device
(analyzer) from the master.
1.
With the “GLI SETUP” top-level menu screen dis-
played, select the “HART INFO” line and press
key
.
2.
With the “HART INFO” submenu screen displayed, se-
lect the “Poll addr” line and press the
key
to display
its related screen.
3.
Assign a polling address of “0” for one analyzer in a
point-to-point configuration, or 1 through “XX” for two or
more analyzers in a multi-drop configuration. Use the
alphanumeric keys to directly select the number, or the
arrow keys to adjust the number digit by digit.
4.
Press the
F4 key
to enter the polling address, and the
F2 key
to send the polling address to the analyzer.
The “Num req preams” information screen shows the num-
ber of preambles required by the analyzer from the master.
1.
With the “GLI SETUP” top-level menu screen dis-
played, select the “HART INFO” line and press
key
.