PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 5-1000
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
61
Table C -- ANALYZER CONFIGURATION SETTINGS (RANGES/CHOICES and DEFAULTS -- continued)
Displayed Screen Title
Entry Range or Choices (where applicable)
Factory Default
Your Setting
RELAY Configuration Settings (continued)
Settings for Only ALARM Relays (continued):
HIGH ALARM?
CONDUCTIVITY:
µS/cm:
0-2.000, 0-20.00, 0-200.0, or 0-2000
mS/cm: 0-2.000, 0-20.00, 0-200.0, or 0-2000
RESIST: 0-19.99 M
Ω
•
cm or 0-999.9 K
Ω
•
cm
TDS: 0-9999 ppm or 0-9999 ppb
TEMP: -20.0 to +200.0
°
C or -4.0 to 392.0
°
F
CONDUCTIVITY:
µS/cm: 2.000, 20.00, 200.0,
or 2000
mS/cm: 2.000, 20.00, 200.0,
or 2000
RESIST: 19.99 M
Ω
or 999.9 K
Ω
•
cm
TDS: 9999 ppm or 9999 ppb
TEMP: 200.0
°
C or 392.0
°
F
LOW DEADBAND?
CONDUCTIVITY: 0-10% of range
RESISTIVITY: 0-10% of range
TDS: 0-10% of range
TEMPERATURE: 0-10% of range
COND: 0 µS/cm or 0 mS/cm
RESIST: 0 M
Ω
•
cm or 0 K
Ω
•
cm
TDS: 0 ppm or 0 ppb
TEMP: 0.0
°
C or 0.0
°
F
HIGH DEADBAND?
CONDUCTIVITY: 0-10% of range
RESISTIVITY: 0-10% of range
TDS: 0-10% of range
TEMPERATURE: 0-10% of range
COND: 0 µS/cm or 0 mS/cm
RESIST: 0 M
Ω
•
cm or 0 K
Ω
•
cm
TDS: 0 ppm or 0 ppb
TEMP: 0.0
°
C or 0.0
°
F
Settings for Only CONTROL Relays:
PHASE?
HIGH or LOW
For Relays A and B: HIGH
SET SETPOINT?
CONDUCTIVITY:
µS/cm:
0-2.000, 0-20.00, 0-200.0, or 0-2000
mS/cm: 0-2.000, 0-20.00, 0-200.0, or 0-2000
RESIST: 0-19.99 M
Ω
•
cm or 0-999.9 K
Ω
•
cm
TDS: 0-9999 ppm or 0-9999 ppb
TEMP: -20.0 to +200.0
°
C or -4.0 to 392.0
°
F
CONDUCTIVITY:
µS/cm: 2.000, 20.00, 200.0,
or 2000
mS/cm: 2.000, 20.00, 200.0,
or 2000
RESIST: 19.99 M
Ω
or 999.9 K
Ω
•
cm
TDS: 9999 ppm or 9999 ppb
TEMP: 200.0
°
C or 392.0
°
F
DEADBAND?
CONDUCTIVITY: 0-10% of range
RESISTIVITY: 0-10% of range
TDS: 0-10% of range
TEMPERATURE: 0-10% of range
COND: 0 µS/cm or 0 mS/cm
RESIST: 0 M
Ω
•
cm or 0 K
Ω
•
cm
TDS: 0 ppm or 0 ppb
TEMP: 0.0
°
C or 0.0
°
F
OVERFEED TIMER?
0-999.9 minutes
0 minutes
PASSCODE Configuration Setting
SET PASSCODE?
DISABLED or ENABLED
DISABLED
TEST/MAINTENANCE Simulation Function Settings
SELECT SIM?
SENSOR A, SENSOR B, TEMPERATURE
A, TEMPERATURE B or CALCULATED
SENSOR A
SIM SENSOR?
Same ranges as those listed for
“SET 4 mA VALUE” function in the
“OUTPUT Configuration Settings” category.
Present measured value of
Sensor A’s selected parameter.