PART THREE - OPERATION
SECTION 5 - ANALYZER CALIBRATION
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
68
☞
Table D -- CONDUCTIVITY REFERENCE SOLUTIONS
Desired Solution Value
Grams NaCl
µS/cm
mS/cm
ppm (NaCl)*
To Be Added
100
200
500
0.10
0.20
0.50
50
100
250
0.05
0.10
0.25
1000
2000
3000
1.00
2.00
3.00
500
1010
1530
0.50
1.01
1.53
4000
5000
8000
4.00
5.00
8.00
2060
2610
4340
2.06
2.61
4.34
10,000
20,000
10.00
20.00
5560
11,590
5.56
11.59
*When using ppm measuring scale for compounds other than
NaCl, refer to appropriate chemistry handbook for reference
solution formulation.
B. Thoroughly rinse the clean sensor in de-
ionized water. Then fully immerse the sensor
electrodes in the prepared reference solution.
Important: Allow the sensor and solution
temperatures to equalize.
Depending on their
temperature differences, this may take a few
minutes.
NOTE:
Suspend the sensor to prevent its
electrodes from touching the container.
Remove any bubbles clinging to the
outer and center electrodes by gently
stirring the sensor in the solution.
Models with a deeper recessed center
electrode require more vigorous stirring.
2.
Press
MENU key
to display
&21),*85(
7(670$,17
(;,7
.
3.
With the “CALIBRATE” line selected, press
ENTER key
to display
6(1625 %
2873876
(;,7
.
4.
With the “SENSOR A” line selected, press
ENTER key
to display
.