PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 5-1000
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
59
4.8 Configuration Setting
Summary
Table C lists all configuration settings and their entry ranges/
choices and factory defaults, categorized by basic functions.
Table C -- ANALYZER CONFIGURATION SETTINGS (RANGES/CHOICES and DEFAULTS)
Displayed Screen Title
Entry Range or Choices (where applicable)
Factory Default
Your Setting
LANGUAGE Configuration Setting
LANGUAGE?
ENGLISH, FRENCH, GERMAN, SPANISH, etc.
ENGLISH
SENSOR Configuration Settings (A and B)
SELECT MEASURE?
CONDUCTIVITY, RESISTIVITY or TDS
CONDUCTIVITY
SENSOR:
DISPLAY FORMAT?
(full scale value)
CONDUCTIVITY:
µS/cm: 2.000, 20.00, 200.0, or 2000
mS/cm: 2.000, 20.00, 200.0, or 2000
RESISTIVITY: XX.XX M
Ω
•
cm or XXX.X K
Ω
•
cm
TDS: XXXX ppm or XXXX ppb
CONDUCTIVITY:
For 0.05 cell: 20.00 µS/cm
For 0.5 cell:
200.0 µS/cm
For 1.0 cell:
2000 µS/cm
For 5.0 cell:
20.00 mS/cm
For 10 cell:
200.0 mS/cm
RESISTIVITY: XX.XX M
Ω
•
cm
TDS: XXXX ppm
SENSOR:
T-COMPENSATION?
LINEAR, AMMONIA, NATURAL WATER,
OPTIONAL TABLE or NONE
Sensor A: LINEAR at 2.00%
per
°
C with 25.0
°
C
reference temp.
Sensor B: MANUAL
TDS: SELECT FACTOR?
NaCl or USER DEFINED
NaCl
TDS: SET FACTOR?
0.01-99.99 ppm/µS
0.49 ppm/µS
LINEAR: SET SLOPE?
0-4.00% per
°
C
2.00% per
°
C
LINEAR: SET REF TEMP?
0-200.0
°
C or 32-392
°
F
25.0
°
C or 77
°
F
SELECT CELL K?
0.05, 0.5, 1.0, 5.0 or 10
1.0
SET CELL K?
0.0500-10.0000
1.0000
SET FILTER?
0-60 seconds
0 seconds
PULSE SUPPRESS?
OFF or ON
OFF
ENTER NOTE?
Enter up to eight characters to replace
COND A or COND B
Sensor A: COND A
Sensor B: COND B
TEMP ELE: SELECT TYPE? PT1000, PT100 or MANUAL
PT1000
TEMP ELEMENT:
SET T FACTOR?
950-1050 ohms or 95-105 ohms
1000 OHMS
TEMP ELE: SET MANUAL?
0.0-200.0
°
C
25.0
°
C
CALCULATION Configuration Settings (Sensor A and B)
CALCULATION:
SELECT MEASURE?
NONE, % REJECT, % PASS, RATIO (A/B),
RATIO (B/A), DIFF (A-B) or DIFF (B-A)
NONE
CALCULATION:
DISPLAY FORMAT?
% REJECT: non-selectable
% PASS: non-selectable
RATIO: XXXX, XXX.X, XX.XX, or X.XXX
DIFF: Same ranges as those listed in
“Sensor A and B Configuration
Settings” category above
% REJECT: 0-100%
% PASS: 0-100%
RATIO A/B or B/A: XXXX
DIFF: Same defaults as those listed
in the “Sensor A and Sensor B Con-
figuration Settings” category above
Not applicable
Not applicable
(Table C continued on next page.)