Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
8
CONDENSED OPERATING INSTRUCTIONS
D. CALIBRATING THE ANALYZER -- (continued)
8.
After the
screen re-appears, press
ESC key
once to display
',63/$< )250$7
&203(16$7,21
&21),* &21'8&7
&21),* /,1($5
&(// &2167$17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
.
9.
Use
Ø
Ø
key
to select the “TEMP ELEMENT” line, and press
ENTER key
to display
6(7 7 )$&725
6(7 0$18$/
(;,7
.
10. Use
Ø
Ø
key
to select the “SET T FACTOR” line, and press
ENTER key
to display a “T FACTOR”
value screen like
. Adjust the displayed value to exactly match the sensor’s
GLI-certified “T FACTOR.” (Use
Ö
Ö
and
Õ
Õ
keys
to coarse adjust, and
×
×
and
Ø
Ø
keys
to fine ad-
just.) Then press
ENTER key
to enter the value.
11. After the
6(7 7 )$&725
6(7 0$18$/
(;,7
screen re-appears, press
MENU key
once and then
ESC key
once
to display the MEASURE screen.
This completes GLI’s DRY-CAL calibration. The analyzer is now ready to measure conductivity.
To change the display format of the MEASURE screen (for example, from 0-2000 µS/cm to 0.000-2.000
mS/cm), refer to PART THREE, Section 4.2, subheading “Select DISPLAY FORMAT.”
NOTE:
If the values you intend to measure are above the analyzer’s set measuring range (not its
selected display format), use a different sensor that has the appropriate nominal cell con-
stant. For a listing of sensor cell constants and their inherent measuring ranges, see Table A
in PART THREE, Section 4.2, subheading “Select DISPLAY FORMAT.”
E. COMPLETING ANALYZER CONFIGURATION
To further configure the analyzer to your application requirements, use the appropriate CONFIGURE
screens to make selections and “key in” values. Refer to PART THREE, Section 4 for complete con-
figuration details.