PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
38
4.2 Configuring
Sensor (A and B)
Characteristics
☞
SELECT MEASURE
(conductivity,
resistivity or TDS)
The analyzer can be used with one or two sensors, provid-
ing up to two separate measurement channels. Each sensor
can be independently selected for different measurements
(conductivity, resistivity or TDS). Also, each sensor can
have a different nominal cell constant.
The analyzer must be configured to define the characteris-
tics of each sensor including its “K” value, temperature
element type, “T” factor, and other related items such as
temperature compensation, input signal filtering, pulse sup-
pression, etc.
When using two sensors, configure them
in the same way using their respective menu screens.
NOTE:
For calculated Sensor A and B measurements (%
reject, % pass, ratio A/B, ratio B/A, diff. A-B or diff.
B-A), both sensors must be configured for the same
measurement (conductivity, resistivity or TDS).
However, each sensor may be set for a different
display format and a different cell constant.
1.
With the
6(7 287387
6(7 5(/$< $
6(7 5(/$< %
6(7 5(/$< &
6(7 5(/$< '
6(7 3$66&2'(
6(7 & 25 )
/$1*8$*(
6(1625 $
6(1625 %
&$/&8/$7,21
(;,7
screen displayed, use
Ø
Ø
key
to select “SENSOR A” line.
2.
Press
ENTER key
to display
',63/$< )250$7
&203(16$7,21
&21),* &21'8&7
&21),* /,1($5
&(// &2167$17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
.
3.
With the “SELECT MEASURE” line selected, press
ENTER key
to display
. Use
Ø
Ø
and
×
×
keys
to view the three choices:
•
CONDUCTIVITY:
Selects conductivity measurement.
•
RESISTIVITY:
Selects resistivity measurement.
•
TDS:
Selects total dissolved solids measurement.
WARNING:
WHEN THE MEASUREMENT IS CHANGED, ALL
USER-ENTERED VALUES ARE REPLACED BY
FACTORY-DEFAULT VALUES.