PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 5-1000
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
43
SELECT CELL K
(sensor’s GLI-certified
“K” Value)
6.
After the
6(7 5() 7(03
(;,7
screen re-appears, press
ESC
key
once to return to “SENSOR A” submenu screen.
Each GLI sensor has a unique, certified “K” value shown on a
label attached to the sensor cable or to the inside cover of its
optional junction box head. By entering this “K value,” calibra-
tion (including zeroing) is only necessary when this sensor is
replaced. This also sets the analyzer measuring range to
match the inherent range of the sensor’s cell constant.
1.
With the
',63/$< )250$7
&203(16$7,21
&21),* &21'8&7
&21),* /,1($5
&(// &2167$17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
screen displayed, use the
Ø
Ø
key
to select the “CELL CONSTANT” line.
2.
Press
ENTER key
to display
.
3.
With the “SELECT CELL K” line selected, press
ENTER
key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to select the nominal cell category that cor-
responds to the sensor’s GLI-certified “K” value, and
press
ENTER key
to enter this selection.
4.
After the
screen re-appears, use
Ø
Ø
key
to select the “SET CELL K” line.
5.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the dis-
played value to exactly match the sensor’s GLI-certified
“K” value, and press
ENTER key
to enter the value.
6.
After the
screen re-appears, press
ESC
key
once to return to “SENSOR A” submenu screen.