PART TWO - INSTALLATION
SECTION 1 - UNPACKING
Rev. 5-1000
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
21
P A R T T W O - I N S T A L L A T I O N
S
ECTION
1
After unpacking, it is recommended to save the shipping
carton and packing materials in case the instrument must be
stored or re-shipped. Inspect the equipment and packing
materials for signs of shipping damage. If there is any evi-
dence of damage, notify the transit carrier immediately.
S
ECTION
2
2.1 Location
2.2 Mounting
1. It is recommended to locate the analyzer as close as
possible to the installed sensor. The maximum allow-
able distance between an installed sensor and the
analyzer is 300 feet (91 m).
2. Mount the analyzer in a location that is:
➥
Clean and dry where there is little or no vibration.
➥
Protected from corrosive fluids.
➥
Within ambient temperature limits (-4 to +140
°
F or
-20 to +60
°
C).
CAUTION:
EXPOSING THE ANALYZER TO DIRECT
SUNLIGHT MAY INCREASE THE OPERATING
TEMPERATURE ABOVE ITS SPECIFIED
LIMIT, AND DECREASE DISPLAY VISIBILITY.
RECOMMENDATION: IN SEVERE CASES,
USE A GLI SUN SHIELD (P/N 1000G3088-001).
Figure 2-1 illustrates the various ways to mount the ana-
lyzer using the supplied bracket and hardware. Determine
the mounting method and attach the hardware as shown in
the respective illustration. Refer to Figure 2-2 for analyzer
installation dimension details.
UNPACKING
MECHANICAL REQUIREMENTS