PART THREE - OPERATION
SECTION 6 - TEST/MAINTENANCE
Rev. 5-1000
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
77
6.7 SELECT SIM
Measurement
6.8 SIM SENSOR Setting
You can simulate a measured value to make the relays and
analog outputs respond accordingly. First, select the type of
simulated value using this subsection. Then, set the desired
simulation value following the steps in subsection 6.8.
1.
With the
screen displayed, use
Ø
Ø
key
to select the “SELECT SIM” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view all choices:
•
SENSOR A:
Depending on the sensor’s configured
measurement, selects the simulated value to be a
Sensor A conductivity, resistivity or TDS value.
•
SENSOR B:
Depending on the sensor’s configured
measurement, selects the simulated value to be a
Sensor B conductivity, resistivity or TDS value.
•
TEMPERATURE A:
Selects the simulated value to
be a Sensor A temperature value.
•
TEMPERATURE B:
Selects the simulated value to
be a Sensor B temperature value.
•
*CALCULATION:
Selects the simulated value to be a
calculated Sensor A and B measurement value (% re-
ject, % pass, ratio A/B, ratio B/A, diff. A-B or diff. B-A).
*A simulated value can only represent the calculated measure-
ment when two sensors are used and the analyzer has been
correctly configured for CALCULATION.
3.
With the desired choice displayed, press
ENTER key
to enter this selection and return to the TEST/MAINT
top-level menu screen.
After selecting the type of simulated measurement (subsec-
tion 6.7), set the desired simulation value.