PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
42
CONFIG LINEAR
Temperature Compensation
(not needed for other
compensation methods)
☞
Only when LINEAR is the selected temperature com-
pensation, determine if the analyzer should be further
configured by entering a specific slope (% per
°
C), and
a reference temperature.
If the built-in AMMONIA or
NATURAL WATER properties table or NONE was selected,
disregard this subsection -- no compensation configuration
is needed.
NOTE:
If “OPTIONAL TABLE” compensation was selected, the
analyzer must be equipped with a customer-specified,
factory-configured temperature compensation table, or
you must configure a table using the optional GLI soft-
ware tool kit 1000G3311 and an IBM-compatible PC.
LINEAR compensation factory defaults are 2.00%/
°
C slope
and 25.0
°
C reference temperature.
These values are ap-
propriate for most aqueous solutions.
Use chemical
handbook tables to find values for uncommon solutions. To
enter different values, do the following:
1.
With the
',63/$< )250$7
&203(16$7,21
&21),* &21'8&7
&21),* /,1($5
&(// &2167$17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
screen displayed, use the
Ø
Ø
key
to select the “CONFIG LINEAR” line.
2.
Press
ENTER
to display
6(7 5() 7(03
(;,7
.
3.
With the “SET SLOPE” line selected, press
ENTER key
to display a screen like
. Use
arrow
keys
to adjust the displayed value to the desired % per
°
C slope, and press
ENTER key
to enter the value.
4.
After the
6(7 5() 7(03
(;,7
screen re-appears, use
Ø
Ø
key
to select the “SET REF TEMP” line.
5.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the dis-
played value to the desired reference temperature, and
press
ENTER key
to enter it.