Troubleshooting
Instruction manual
46
Battery errors
discharged state and the instrument is about to store the
internal data and shut-down.
Response: Recharge the battery.
Set-up errors
Pipe OD out of range
Interpretation: You have entered an out-of-range value
for the pipe outside diameter dimension – i.e. larger or
smaller than the unit or sensor can be used on.
Response: Enter a valid number.
Wall thickness out of
range
Interpretation: You have entered an out-of-range value
for the pipe wall thickness dimension – accepted range is
1mm - 75mm.
Response: Enter a valid number.
Lining thickness out of
range
Interpretation: You have entered an out-of-range value
for the lining thickness dimension – acceptable range is
0mm - 25mm.
Response: Enter a valid number.
Temperature range
Interpretation: You have entered an out-of-range value
for the fluid Temperature. Accepted temperature range -
20°C to +300°C.
Response: Enter a valid number.
Invalid Date or Time
Interpretation: The entered Date or Time is invalid, or
when setting up ‘timed’ data logging the Stop time is set
earlier than the Start time.
Response: Enter a valid Date and Time.
Sensors: INVALID
Interpretation: The selected temperature is higher than
the maximum allowed for the sensor type.
Response: Select alternative sensors or change the
temperature.
Mode: Err Typ
Interpretation: The selected sensors are invalid and the
mode cannot be verified.
Response: Select a valid sensor type and choose a mode
that gives a non-zero separation distance.