Instrction manual
Operating Procedures
23
1.
Once Select View Edit Site Data from the MAIN
MENU.
2.
Select Choose from list of sites.
3.
Select one of the sites listed and press the
ENTER key.
4.
The Site name will show the selected site and
the site parameters will be listed on the
screen.
5.
Scroll down through the menu list and
enter/change the data that might have
changed since the last time the site was
accessed.
6.
When you are satisfied that the parameters
are correct select Save current site & read
flow.
If you select a different Sensor set (e.g. A-ST) when entering new site data you could receive
an “invalid” error message if the previous Sensor set was operating at a temperature above
135ºC. If this occurs, ignore the warning as it will disappear when you enter a temperature in
the correct range for new sensors.
7.
If you need to alter the fluid temperature from
that shown select either °C or °F with the
cursor and press the ENTER key.
8.
Enter the new temperature value and press
the ENTER key.
9.
The new temperature should now be indicated
in both °C and °F.
10.
Select Continue.. and then press the ENTER
key.
11.
The SENSOR SEPARATION screen now
displays a summary of the entered parameters
and informs you of the type of sensor to be
used, the mode of operation and the distance
to set up between the sensors. In this example
it recommends type A-ST (A standard) sensors
operating in the ‘Reflex’ mode spaced at
32.2mm apart.
12.
Take a note of the displayed parameters, then
press the ENTER key.
13.
The ATTACH SENSORS screen displays, giving
instructions to attach the sensors.
NOTICE