Operating Procedures
Instruction manual
34
item. The log must be cleared before new times
can be entered.
5.
Select Stop date & time and enter the date and
time you wish logging to cease. Note this must
be later than the start time.
6.
Select Set auto start. This enables the auto
logging application.
If Set Auto start is enabled with valid start and
stop times then this entry will change to
Cancel Auto start.
7.
To cancel the automatic logging session
before it commences, click the Cancel Auto
Start.
8.
Select Exit to return to the FLOW READING
screen.
Monitoring the logged events
9.
If you wish to view the logging progress at any time while you are operating in FLOW
READING mode, press the Logger function key as described above to access the REAL
TIME LOGGER screen. You can then select View log as text or View log as graph, as
required, without interfering with the logging operation.
Manually stopping the auto logging session
10.
The logging session will stop automatically at
the programmed Stop date & time.
11.
If you wish to manually stop logging earlier
than programmed, press the Logger function
key to access the REAL TIME LOGGER screen
then select STOP NOW to cease logging.
12.
Select Exit to return to the FLOW READING
screen.
The logged data will remain stored in the
instrument’s memory and can be viewed at any
time as described above.
NOTICE
NOTICE