Operating Procedures
25
and cannot be renamed; subsequent sites are initially named EmptySite1 through to
EmptySite19.
7.4.1
Setting up a new site
Once you have set up a named site you are advised to log some data to the site in order to
avoid the possibility of the site name being overwritten when the QuickStart data is saved.
1.
Select View Edit Site Data from the MAIN
MENU.
2.
Select Choose from list of sites.
3.
Select one of the EmptySites from the
presented list (e.g. EmptySite 1 as shown).
4.
Select Site name and press the ENTER key.
5.
This opens the ALTER NAME screen.
6.
Select Alter the Site Name and you will be
presented with a screen which allows you to
enter a new name in much the same way as
when composing a mobile text message.
7.
On completion press the ENTER key then
select Exit. This will take you back to the VIEW
EDIT SITE DATA screen.
8.
Scroll down through the menu list and enter/
change the pipe parameters and other data
pertaining to the site. Note that this menu
allows you to choose a Sensor Set, unlike the
QUICK START menu which recommended the
sensor set to use. If you enter an inappropriate
sensor set in this menu you will be presented
with an error message later when you go to
the SENSOR SEPARATION screen.
9.
When all the data is correct you can either:
a) Select Save current site & read flow to
continue fitting the transducers and
opening the FLOW READING screen.
b) Select Delete this site to delete the site
name and values and restore it to the
original EmptySite name.
c) Select Exit to return to the MAIN MENU.