Operating Procedures
Instruction manual
20
Entering the site data
1.
Select Quick Start from the MAIN MENU and
press the ENTER key. You will then be
presented with a series of screens in which to
enter the data mentioned above.
2.
Select the dimension units (millimetres or
inches) used to measure the pipe, then press
the ENTER key.
3.
Enter the pipe outside diameter dimension,
then press the ENTER key.
4.
Enter the pipe wall thickness dimension, then
press the ENTER key.
5.
If the pipe has a lining, enter the lining
thickness. If nothing is entered the instrument
automatically assumes there is no lining.
6.
Press the ENTER key to continue
7.
Select the pipe wall material from the list
provided, then press the ENTER key.
If the material is not listed select Other and
enter the propagation rate of the pipe wall
material in metres/sec. Contact your local GF
partner if this is not known.
8.
If a lining thickness value was entered earlier,
this screen is displayed to request that you
enter the lining material type. If no lining
thickness was entered this screen will be
bypassed.
9.
Select the lining material from the list
provided then press the ENTER key. If the
material is not listed select Other and enter
the propagation rate of the lining material in
metres/sec. Contact your local GF partner if
this is not known.
10.
Select the fluid type from the list provided and
press the ENTER key. If the liquid is not listed
select Other and enter a propagation rate in
metres/second.