DESCRIPTION
Cold tap water is automatically injected into the drain discharge to lower the
discharge temperature before entering into the building's drain system.
USER CONTROLS
None
MAINTENANCE REQUIRED
The drain discharge temperature must be checked with a calibrated
thermometer to ensure the proper amount of cold water is being injected
into the drain.
Steam and Water Pressure
Gauges
Provided on the incoming steam and water line are pressure gauges to
monitor the facility utilities.
Vapor Removal Condenser
A condenser is provided to reduce excessive vapor and cool the exhaust air
prior to entering the building exhaust system. The exhaust vapor is reduced.