Series 1800 Washers
61301604086
7–11
During programming mode, select the agent to be injected during the
wash 1 and 2 phase. On text controls, use the TOGGLE button to change
from Acid to Alkaline.
Figure 7–9. TOGGLE BUTTON BEFORE
Manual mode causes the Alkaline detergent pump to be activated for a
preset period of time.
Figure 7–10. TOGGLE BUTTON AFTER
Concentration must be tested on a monthly basis to ensure proper amount
of chemical is being injected into Wash 1 and Wash 2 sump.
PUMP ON/OFF INDICATOR
Red light that comes on when the pump is energized (automatically or
manually).
PUMP
Draws liquid from the supply source and feeds it to the wash sump.
ACID
DETERGENT
ALKALINE
DETERGENT
NOTE