Series 1800 Washers
61301604086
4–5
Six (6) Months
LUBRICATE THE PUMP MOTOR
Re-lubricate the sleeve with a high grade ball and roller bearing grease
every six (6) to twelve (12) months as indicated on the motor.
CHECK THE GFI SENSOR
ELECTRIC SHOCK HAZARD: Only an electrician should
perform this test!
1.
Turn on power to the washer.
2.
Press test button on the Ground Fault Interrupt (GFI) located on
back panel.
3.
Power to the unit should go off.
4.
Press reset.
5.
Power to the unit should come back on.
6.
Fill out GFI chart on the inside of the electrical box.
If this test fails, contact your Authorized Service Technician.
WARNING
WARNING