Troubleshooting
5–4
If this alarm appears on the screen, the thermo-couple wire is broken or
disconnected.
1.
Trace out the thermo-couple wire (Type K) from the control
panel to the sump temperature probe.
2.
Repair or replace as needed.
If this alarm appears on the screen, the thermo-couple wire is broken or
disconnected.
1.
Trace out the thermo-couple wire (Type K) from the control
panel to the sump fill temperature probe.
2.
Repair or replace as needed.
If this alarm appears on the screen, the “CYCLE ABORT” button has been
pressed.
1.
Press “ANSWER ALARM” (cycle will have to be started
from the beginning).