The Control Panel
2–30
FillTemperature Settings
1.
Turn water to full open setting
2.
Turn steam to full open setting (temperature not to exceed 195
o
F).
3.
If water at full open and steam at full open does not create an out put
temperature of 195
o
F, turn water flow down until 195
o
F is achieved.
4.
If water at full open and steam at full open exceeds 195
o
F, turn steam
throttle valve down until in 195 range.
5.
If temperature becomes unstable turn the water pressure regulator
down to 5 psi lower than dynamic house pressure and repeat steps
1 to 5.
STEAM FLOW
SUMP FILL
STEAM VALVE
STEAM
THROTTLING
VAVLE
SUMP FILL
VALVE
STEAM
TRAP
VACUUM
BREAKER
WATER
THROTTLING
VAVLE
SUMP FILL
TEMPERATURE
GAUGE