Series 1800 Washers
61301604086
5–3
Screens Activated by Alarms
If any of the following messages appear during the cleaning cycle, follow the
instructions as noted.
If this alarm appears on the screen, the sump has taken too long to fill.
1.
Press “ALARM ANSWER”.
2.
Check house water pressure (must be 35PSI minimum)
3.
Check sup0ply valve (valve must be fuly open).
4.
Check air pressure (air pressure must be 80 PSI
minimum).
If this alarm appears on the screen, the sump has taken too long to heat.
1.
Press “ANSWER ALARM”.
2.
Check house steam pressure (must be 30PSI minimum
unless ordered with low pressure option [10 PSI
minimum]).
3.
Check condensate valve (must be fully open).
4.
Check supply valve (must be fully open).
5.
Check air pressure (air pressure must be 80 PSI
minimum).
6.
Check exhaust damper (must go to closed position when
treatment pump is running).