TS4000H
32
En
Energize relay
nL
Non-latching relay
SO2
Sulfur Dioxide (range setting: 20 ppm and 100 ppm)
20
20 ppm scale
100
100 ppm scale
LA
Latching relay
##
Relay set point
Hi
Alarm relay setting (Warning relay for O
2
)
##
Relay set point
Ch1
User Modbus channel 1 settings
Ch2
User Modbus channel 2 settings
Br
Baud rate settings
24
2,400 baud rate
48
4,800 baud rate
96
9,600 baud rate
192
19,200 baud rate
For
Format Settings
8n1
8 bits, no parity, 1 stop bit
8n2
8 bits, no parity, 2 stop bits
8o1
8 bits, odd parity, 1 stop bit
8E1
8 bits, even parity, 1 stop bit
Add
Address Settings
###
Number in range 1 – 247 incrementing by magnet
or
Over range
Hrt
HART settings, HART chip present on the device
on
HART option, enable HART communication
off
HART option, disable HART communication
Table 15: User Menu Display
4.4 Start-up
Upon power-up, the
TS4000H
will display “
8.8.8.
” on three 7-segment LCDs, two Alarm and
Warning LEDs. The software revision letters “
rN
’’ (N – revision letter) are briefly displayed. The
TS4000H then enters Start-up Mode ‘’
SU
”, allowing the electrochemical cell to stabilize. Upon
sensor stabilization, the TS4000H enters Operation Mode and displays the current gas
concentration at the electrochemical cell. For detailed information on the Gas Check and
Calibration Modes, refer to Sections 4.8 and 4.9, respectively.
NOTE:
During power-up, the Base Unit may briefly display “F1” after displaying the software
revision letter.
NOTE:
A new sensor may take up to sixty minutes to stabilize once installed in the TS4000H.